Eye protection for curing lights and dental lasers: blue light peaks at 440 nm, the daily limit can be spent in seconds, and laser eyewear has to name a wavelength and an optical density

Blue-light retinal injury runs from 380 to 550 nm and peaks near 440 nm, which is where curing lights are built to emit. Measured against the ACGIH limit, the whole day's permissible exposure to one plasma arc unit at 30 cm was about 6 seconds. Laser eyewear is a different calculation: optical density at a named wavelength.

Regulation, safety and payers

Issued by

ADA Science and Research Institute and OSHA

2023

Status
current
Source
ada.org
Last read
2026-09-13
Three panels across the top give the curing light numbers: blue-light retinal injury occurs mainly between 380 and 550 nanometers with sensitivity peaking near 440; the ACGIH threshold limit value for an 8-hour day is 10 millijoules per square centimeter for blue light and 3 for ultraviolet; and the shortest measured permissible cumulative daily exposure was about 6 seconds, for a plasma arc unit curing from the palatal aspect with the eye 30 centimeters away. Below, a five-row table of minimum optical densities for laser protective eyewear for direct intrabeam viewing at 10 seconds, 600 seconds and 8 hours: argon at 0.514 micrometers, 3.4, 5.2 and 6.4; helium neon at 0.633 micrometers, 1.1, 1.7 and 2.9; a gallium arsenide diode at 0.840 micrometers, 1.8, 2.3 and 3.7; neodymium YAG at 1.064 micrometers, 4.7, 5.2 and 5.2; and carbon dioxide at 10.6 micrometers, 6.2, 8.0 and 9.7.
The blue light numbers and the laser numbers on one page, because they are not the same kind of number. Blue light from the ADA's curing lights topic and Labrie 2011; optical densities reproduced from OSHA Technical Manual Table III:6-8, which used ANSI Z136.1 maximum permissible exposures. Illustration, not a specification for your equipment.Open full size
where retinal sensitivity to blue light peaks; injury occurs mainly across 380 to 550 nm, the band curing lights are built for
440 nm
ACGIH threshold limit value for blue light over an 8-hour workday, with 3 mJ/cm2 for ultraviolet
10 mJ/cm2
the whole day's permissible cumulative exposure with a plasma arc unit curing from the palatal, eye at 30 cm
About 6 seconds
minimum optical density OSHA's table gives for 600-second intrabeam viewing of a 100 W Nd:YAG at 1.064 um
OD 5.2

Covered

4
  • Anyone in the operatory while a curing light runs, including the patient and an assistant at 50 cm; the ADA's rule of thumb is that patient eyewear goes on first and comes off last
  • The operator of any dental laser, plus everyone else in the room, since ANSI Z136.1 control measures as OSHA describes them turn on the class of the device rather than on who is holding it
  • Prescription glasses wearers, who are not covered by their own lenses: ordinary prescription glasses do not prevent penetration of blue or UV light, and contact lenses are no barrier at all
  • Loupes users, whose pupil-level blue irradiance was measured up to eight times higher with loupes than without, though the enlarged retinal image lengthened the permissible daily time rather than shortening it

Not covered

4
  • Ultraviolet, at least from the four curing units measured. The maximum permissible daily exposure for UV exceeded 8 hours at all distances and orientations, and the authors concluded those units did not pose a risk of UV-mediated ocular damage
  • Federal OSHA numbers for blue light: there are none. OSHA has no comprehensive laser standard either, and cites ANSI Z136.1 through the general duty clause and the personal protective equipment rules instead
  • Anything requiring a specific optical density at erbium wavelengths. OSHA's table covers argon, helium neon, gallium arsenide, Nd:YAG and CO2; it does not publish values near 2,780 or 2,940 nm, and ANSI Z136.1 and Z136.3 are sold, not free, so no figure for those is quoted here
  • The clinical question of whether a laser helps, which is a separate body of evidence. See slug aap-best-evidence-consensus-lasers-2018

Timeline

  1. 1986

    1. 1986

      The ADA recommends protective filtering eyeglasses when using quartz-tungsten-halogen curing units

  2. 1993

    1. 1993

      The ANSI Z136.1 edition whose maximum permissible exposures underlie the optical density table in OSHA's Technical Manual

  3. 2007

    1. Jun

      Nordic Institute of Dental Materials tests 18 protective filters; 9 pass, transmitting under 0.1 percent between 400 and 525 nm

  4. 2011

    1. 2011

      Labrie and colleagues publish measured maximum permissible daily ocular exposure times for four curing light types in JCDA

  5. 2013

    1. 2013

      ISO 12609-1 sets transmittance requirements and a B-1 to B-6 blue-light classification for eyewear against intense light sources in medical and cosmetic use

  6. 2016

    1. Aug

      Price and colleagues measure the effect of magnification loupes on the blue light hazard: about 11 minutes a day at 40 cm without loupes, up to 28 with them

  7. 2019

    1. Dec

      Fluent and colleagues note in JADA that the CDC dental infection control guidelines and the OSHA bloodborne pathogens standard say nothing about blue light

  8. 2023

    1. Mar 27Latest

      The ADA's dental curing lights topic page, last updated this date, reports that 9 of 22 filtering devices tested in the ADA laboratory transmitted blue light above 4 to above 15 percent

What it says

There are two optical hazards in a dental operatory and they behave differently, which is why one pair of glasses does not answer both.

The curing light hazard is photochemical and cumulative. The ADA's dental curing lights topic page states that blue-light retinal injury takes place primarily from exposure in the wavelength range between 380 and 550 nm, with the sensitivity of the retina peaking at approximately 440 nm, and that because camphorquinone absorbs from about 455 to 481 nm with a peak near 469 nm, dental curing units are optimized to perform in exactly the wavelength range where the injury happens. The page reports that under certain clinically relevant conditions the light from dental curing units may exceed dose limit values for photochemical retinal exposure over an 8-hour workday after just under three hours of exposure.

The measured version of that arithmetic is in Labrie and colleagues, published free in the Journal of the Canadian Dental Association in 2011. They cured a simulated incisor restoration from the facial and the palatal with four unit types, plasma arc, low-power LED, high-power LED and quartz-tungsten-halogen, measuring spectral irradiance at 30, 50 and 100 cm, standing in for the operator, the assistant and an onlooker. Weighted against the ACGIH threshold limit values, which in an 8-hour workday are 10 mJ/cm2 for blue light and 3 mJ/cm2 for UV, the maximum permissible cumulative daily exposure to blue light ran from as low as 5.96 seconds, curing from the palatal with the plasma arc unit at 30 cm, to as high as 1.5 hours with the low-power LED from the facial. Put in cycles, an operator 30 cm from a palatally positioned light guide spends the whole day's allowance after two 5-second cycles with that plasma arc unit, two 20-second cycles with the high-power LED, or four 20-second cycles with the low-power LED or the halogen unit. UV was not the problem: the permissible daily UV exposure exceeded 8 hours at every distance and orientation, and the authors concluded the four units tested did not pose a risk of UV-mediated ocular damage.

Distance and loupes change the number. Price and colleagues measured light reflected off a tooth at 40 cm and calculated a permissible cumulative daily exposure of about 11 minutes without loupes; with loupes the irradiance at the pupil was up to eight times greater, but because retinal image area grows with the square of the magnification the effective hazard fell and the permissible time rose to as much as 28 minutes.

The protection is where the evidence turns uncomfortable. Labrie's group notes that ordinary prescription glasses do not prevent penetration of blue or UV light, and that some glasses filtering blue and UV reduce transmission below 500 nm to less than 1 percent; with appropriate eye protection, the same operator could look at the plasma arc unit at 30 cm for a total of 10 minutes a day. Bruzell and colleagues tested 18 protective filters and found nine adequate, those nine transmitting less than 0.1 percent of radiation at any wavelength between 400 and 525 nm. The ADA's own laboratory found 9 of 22 protective filtering devices transmitting blue light from dental curing units at levels ranging from above 4 percent to above 15 percent under clinically relevant conditions. And the ADA states plainly that there is currently no standard that specifically sets test methods, requirements and labeling for protective filtering devices intended for protection against retinal blue light exposure from dental curing units; a working draft sits in a subcommittee of ISO Technical Committee 106, and in the meantime ISO 12609-1 covers eyewear for intense light sources in cosmetic and medical settings with a B-1 to B-6 blue-light classification.

Lasers are a thermal and photochemical hazard governed by wavelength and power, and the eyewear is specified by a number. OSHA's Technical Manual chapter on laser hazards explains the framework: manufacturers classify and label under the federal laser product performance standard at 21 CFR Part 1000, with classes defined at 21 CFR 1040.10 as Class I, IIa, II, IIIa, IIIb and IV; ANSI Z136.1 is the user standard supplying maximum permissible exposures; and optical density is the base-ten logarithm of the worst-case exposure divided by the MPE. Because OD is logarithmic, OD 3 transmits a thousandth of the incident energy and OD 6 a millionth. OSHA reproduces a table of minimum optical densities for direct intrabeam viewing computed from ANSI Z136.1 maximum permissible exposures, which includes a 1 W argon at 0.514 um needing OD 3.4 for 10 seconds, 5.2 for 600 seconds and 6.4 for 8 hours; a 5 mW helium neon at 0.633 um at 1.1, 1.7 and 2.9; a 50 mW gallium arsenide diode at 0.840 um at 1.8, 2.3 and 3.7; a 100 W Nd:YAG at 1.064 um at 4.7, 5.2 and 5.2; and a 1000 W CO2 at 10.6 um at 6.2, 8.0 and 9.7. The same chapter notes that eye protection is generally required for Class IIIb and Class IV lasers and that all laser eyewear shall be clearly labeled with OD values and the wavelengths it protects against.

What is not there is a federal rule with a number in it. OSHA's chapter says that at present OSHA does not have a comprehensive laser standard, that the construction rules at 29 CFR 1926.54 and the personal protective equipment subpart may apply, and that citations are issued by invoking the general duty clause or Subpart I, requiring employers to work to industry consensus standards such as ANSI Z136.1. The construction eye-protection rule, 29 CFR 1926.102(b)(2), is the one that spells out the requirement in regulatory language: employees exposed to laser beams shall be furnished suitable laser safety goggles which will protect for the specific wavelength of the laser and be of optical density adequate for the energy involved. For general industry, which is where a dental office sits, the enforceable hook is 29 CFR 1910.133(a)(1): the employer shall ensure that each affected employee uses appropriate eye or face protection when exposed to eye or face hazards from, among other things, potentially injurious light radiation.

From the document

Blue-light retinal injury takes place primarily from exposure in the wavelength range between 380 to 550 nm, with the sensitivity of the retina peaking at approximately 440 nm.
ADA oral health topic: Dental Curing Lights

Who it applies to

Everyone in the room. The curing light exposure times above were calculated at 30 cm for an operator, 50 cm for an assistant and 100 cm for an onlooker, and the assistant's numbers are only a few multiples better than the operator's. The ADA's Eye Safety in the Dental Office page adds the patient: protective eyewear goes on as the patient sits down and comes off as they leave, patients with their own prescription glasses still need it, and contact lenses provide no barrier. It also notes that specific protective eyewear is designed for procedures using lasers, light curing units and bleaching units, and that each manufacturer's product literature governs which is which.

Two limits on this entry are worth saying out loud. First, ANSI Z136.1 and the health care companion Z136.3 are sold documents; nothing in them is quoted here beyond the way OSHA characterizes them in its own free manual. Second, OSHA's optical density table was computed for the laser types in the table, and it does not include the erbium wavelengths near 2,780 and 2,940 nm that some dental lasers use. For those, the number comes from the device manufacturer and the eyewear supplier, on the label, not from anything free that could be verified here.

What changed

The lights got stronger and the guidance did not keep up. The quartz-tungsten-halogen units studied in the 1980s delivered around 300 mW/cm2 and were judged to have little potential to cause ocular injury; the plasma arc and LED units in the 2011 study deliver up to 3,000 mW/cm2 or more, with narrow spectra centered near the retina's most sensitive wavelength. LED units are also less likely to trigger the eye's aversion response, which normally limits an unintended glance to about a quarter of a second.

Meanwhile the regulatory picture stayed empty. Fluent and colleagues wrote in JADA in 2019 that the CDC's dental infection control guidelines and OSHA's bloodborne pathogens standard include no recommendations or regulations directly related to blue light exposure, that other OSHA regulations do require employers to protect employees from potentially injurious light radiation, and that it is not readily evident to employers or dental personnel that those regulations apply. The ADA's 2023 page confirms the gap is still open.

What to do in the practice

0 of 7 done

Ticks are saved in this browser only.

Where the source is

The ADA's Dental Curing Lights topic page, last updated March 27, 2023, is the link under the title and is the best single free summary of the blue light hazard, the filter testing and the standards gap. Labrie and colleagues' 2011 paper is free in full text at the Journal of the Canadian Dental Association and contains the measured exposure times and the cycle counts. The Bruzell filter study and the Price loupes study are behind publisher access; their abstracts, which are public, carry the figures quoted here. OSHA's Technical Manual, Section III Chapter 6, is free and carries the laser classes, the optical density definition and table, and OSHA's own statement that it has no comprehensive laser standard. The federal laser product classes are at 21 CFR 1040.10 on eCFR, and the general-industry eye protection rule is 29 CFR 1910.133. ANSI Z136.1 and Z136.3 are sold by the Laser Institute and are not quoted here.

lasersPPEADAOSHAequipmentcomplianceevidencetraining

This page summarizes a published rule or document for dentists. The document itself governs; read it before you act on it, and check with counsel where a rule touches your state.