Regulation · Infection control
Sterilization cycle parameters: the temperatures, times and pressures a dental office runs, and why the sterilizer's own manual wins
CDC publishes recognized exposure conditions for steam (30 minutes at 250°F gravity, 4 minutes at 270°F prevacuum), dry heat (60 minutes at 340°F) and describes unsaturated chemical vapor, but the 2003 dental guideline is explicit that the sterilizer manufacturer's times, temperatures and parameters are the ones to follow.
Issued by
Centers for Disease Control and Prevention
2008
- Status
- current
- Source
- cdc.gov
- Last read
- 2026-09-13
- gravity displacement steam, wrapped instruments, CDC Table 7
- 30 min at 250°F
- prevacuum steam, wrapped instruments, CDC Table 7
- 4 min at 270°F
- dry heat in a hot air sterilizer, CDC 2008
- 60 min at 340°F
- chemical vapor exposure, timed from 20 psi, Chemiclave manual
- 20 minutes
Covered
3- Every heat-tolerant critical and semicritical instrument in the practice, which CDC says is the majority of patient-care items in dentistry and should therefore be heat sterilized
- All three dental heat methods named in the 2003 guideline: steam under pressure (gravity displacement and prevacuum), dry heat (static-air oven and forced-air rapid heat transfer), and unsaturated chemical vapor
- Every tabletop sterilizer in the building separately, because the cycle menu is set by that model's manufacturer and its FDA clearance, not by a general table
Not covered
3- Heat-sensitive items, which CDC discourages: they must be replaced with heat-tolerant or disposable alternatives where possible, and otherwise processed with an FDA-cleared liquid chemical sterilant or high-level disinfectant on that product's own label conditions
- Liquid chemical sterilization, which CDC says cannot be verified with biological indicators, and bead sterilizers, which FDA has deemed neither safe nor effective
- This entry does not reproduce ANSI/AAMI ST79. It is a paid standard, so nothing here is quoted from it; it appears only as CDC characterizes it
Timeline
2003
Dec 19
CDC's dental infection control guideline publishes the sterilization section and defers cycle parameters to the equipment manufacturer
2008
2008
CDC's Guideline for Disinfection and Sterilization in Healthcare Facilities publishes Table 7 (steam cycle minimums) and Table 8 (flash cycle examples)
2010
2010
ANSI/AAMI ST79-2010 issued, amended A1:2010, A2:2011, A3:2012 and A4:2013; CDC lists it as a source on its dental sterilization pages
2016
Oct
CDC's Summary of Infection Prevention Practices restates that reprocessing follows manufacturer instructions
2024
May 15
CDC's dental sterilization page updated: follow the manufacturer's recommendations for times, temperatures and other operating parameters
JunLatest
CDC's 2008 disinfection and sterilization guideline carries its most recent update stamp
What it says
Start with the sentence that settles every argument. CDC's 2003 dental guideline says heat-tolerant dental instruments are usually sterilized by steam under pressure, dry heat, or unsaturated chemical vapor; that all sterilization should use medical sterilization equipment cleared by FDA; and that the times, temperatures and other operating parameters recommended by the manufacturer of the equipment used should always be followed. CDC's current dental sterilization page, updated May 15, 2024, repeats it and adds one line worth taping to the wall: if the manufacturer's instructions for use are not clear, contact the manufacturer.
With that said, CDC does publish numbers, in the companion Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), which carries a June 2024 update stamp and which the dental pages cite as the detailed reference.
Steam. The guideline states that the two common steam-sterilizing temperatures are 121°C (250°F) and 132°C (270°F), and that the four parameters of steam sterilization are steam, pressure, temperature and time. Pressure, it explains, "serves as a means to obtain the high temperatures necessary to quickly kill microorganisms," which is why a steam cycle is specified by temperature and time rather than by a pressure reading. Table 7, minimum cycle times for steam sterilization cycles, gives, for wrapped instruments: gravity displacement, 30 minutes at 250°F (121°C) or 15 minutes at 270°F (132°C), with 15 to 30 minutes drying; dynamic-air-removal, which is the prevacuum type, 4 minutes at 270°F (132°C) with 20 to 30 minutes drying, and no entry at all at 250°F. Textile packs in a gravity sterilizer take 30 minutes at 250°F or 25 minutes at 270°F. Table 8, flash steam parameters, lists 132°C (270°F) for 3 minutes for nonporous items in either machine, 10 minutes for mixed nonporous and porous items in a gravity sterilizer, and 4 minutes for that mixed load in a prevacuum. Both tables are marked as modified from the Association for the Advancement of Medical Instrumentation.
The 2003 dental guideline explains the difference between the two steam machines. Most tabletop sterilizers in dental practice are gravity displacement, in which unsaturated air is forced out through a vent and trapped air is the failure mode: "errors in packaging items or overloading the sterilizer chamber can result in cool air pockets and items not being sterilized." Prevacuum sterilizers pull a vacuum before the chamber is pressurized, which gives faster steam penetration, and they carry their own duty: they "should be tested periodically for adequate air removal, as recommended by the manufacturer," and a unit that fails should not be used until it is inspected and passes.
Dry heat. The 2008 guideline states that the most common time-temperature relationships for hot air sterilizers are 170°C (340°F) for 60 minutes, 160°C (320°F) for 120 minutes, and 150°C (300°F) for 150 minutes; its dental paragraph gives the shorthand "dry heat (e.g., 320ºF for 2 hours)." Dry heat is for things moist heat damages, such as burs and certain orthodontic instruments, and comes in a static-air oven type, slower and less uniform, and a forced-air or rapid heat transfer type. Dry heat spore tests use Bacillus atrophaeus rather than Geobacillus stearothermophilus, because those spores are more resistant to dry heat.
Unsaturated chemical vapor. CDC describes the process rather than the cycle: it "involves heating a chemical solution of primarily alcohol with 0.23% formaldehyde in a closed pressurized chamber," and it corrodes carbon steel instruments such as dental burs less than steam does because so little water is present. Instruments must be dry going in, and the spent solution is hazardous waste under state and local rules. For the cycle you must read the machine's manual. The Harvey Chemiclave EC5500 and EC6000 owner and operator manual describes a 20-minute timed exposure phase in which the timer does not start until chamber pressure reaches 20 psi, and stops counting if pressure falls below 20 psi; the service manual for the same units sets the chamber temperature acceptance window at 265°F to 275°F. No drying time is required, and the manual rules out linen, textiles, liquids, agars, polycarbonate, nylon bags and anything that cannot tolerate 132°C (270°F).
What a real dental sterilizer offers. Midmark's manual for the M9, M9D and M11, all gravity displacement, lists four preset cycles: Unwrapped, 270°F (132°C) for 3 minutes at 27.1 psi; Pouches, 270°F for 5 minutes at 27.1 psi; Handpieces, 270°F for 6 minutes at 27.1 psi, with a note to verify the parameters with the handpiece manufacturer; and Packs, 250°F (121°C) for 30 minutes at 15 psi. Each carries a 30-minute drying time. Tuttnauer's manual for the EHS 2540 and 3870, pre and post vacuum machines, lists a wrapped cycle at 273°F to 279°F for 8 minutes with 20 minutes drying, a fast wrapped cycle at the same temperature for 4 minutes, a delicate cycle at 250°F to 256°F for 30 minutes, and separate Bowie and Dick air-removal and vacuum leakage test programs. Two machines, both correct, and not one number the same.
From the document
The sterilization times, temperatures, and other operating parameters recommended by the manufacturer of the equipment used, as well as instructions for correct use of containers, wraps, and chemical or biological indicators, should always be followed.
Who it applies to
Any practice that reprocesses instruments, which is all of them. CDC's position is that the majority of patient-care items in dentistry are heat-tolerant and should therefore be heat sterilized, and that includes the semicritical ones: handpieces, low-speed motors and reusable prophy angles are heat sterilized between patients and never high-level or surface disinfected.
The parameters above do not reach heat-sensitive items. CDC discourages them and tells offices to substitute a heat-tolerant or disposable alternative; where none exists, an FDA-cleared liquid chemical sterilant is used on its own label conditions, and CDC notes that process cannot be verified with biological indicators. Bead sterilizers are out: FDA required that commercial distribution cease absent a premarket approval application, and CDC says an office using one assumes the risk of a device FDA has deemed neither safe nor effective.
What changed
The framework has not moved since 2003. What has moved is the paperwork underneath it. CDC's dental sterilization and sterilization monitoring pages, both dated May 15, 2024, now list ANSI/AAMI ST79-2010 with amendments A1:2010 through A4:2013 among their sources, and the 2008 guideline's Tables 7 and 8 are labeled as modified from AAMI. ST79 is a paid standard. I did not read it and nothing here is quoted from it; if your state board or accreditation body cites ST79 directly, buy the copy rather than trust a summary.
The other change is packaging. CDC now asks that every package be labeled with the sterilizer used, the cycle or load number, the date and an expiration date where one applies, so a failed load can be traced and pulled.
What to do in the practice
0 of 7 done
Ticks are saved in this browser only.
Where the source is
The 2008 Guideline for Disinfection and Sterilization in Healthcare Facilities is a free 163-page PDF on cdc.gov; Tables 7 and 8 sit near the back. The 2003 dental guideline is free as an MMWR PDF and as a web page, and its sterilization section runs about pages 22 to 24, which is where CDC's own dental page points readers. Note that cdc.gov refused automated downloads under a browser user agent during this check and answered a plain one, which is a quirk of its edge, not of the documents. The Midmark and Tuttnauer manuals are free PDFs on the manufacturers' own technical libraries and are cited here only as examples of what a real cycle table looks like. The Harvey Chemiclave EC5500 and EC6000 owner and service manuals were read as posted PDFs of the Barnstead International documents; that product line is long discontinued, so treat those numbers as the manual's, not as a current recommendation. ANSI/AAMI ST79 is sold by AAMI and was not read.
infection controlCDCguidelineequipmentcomplianceinspectiontraining
This page summarizes a published rule or document for dentists. The document itself governs; read it before you act on it, and check with counsel where a rule touches your state.